SCANNING ELECTRON MICROSCOPE

SEM ''Cambridge S150'' at Geological Institute, University Kiel, 1980

SEM opened sample chamber

The 'scanning electron microscope' ('SEM') is a type of electron microscope capable of producing high-resolution images of a sample surface. Due to the manner in which the image is created, SEM images have a characteristic three-dimensional appearance and are useful for judging the surface structure of the sample.
The SEM was pioneered by Manfred von Ardenne in the 1930s [1] [2]. The instrument was further developed by Charles Oatley and first commercialized by Cambridge Instruments

Contents
Scanning process
Sample preparation
Detection of secondary electrons
Detection of backscattered electrons
Beam-injection analysis of semiconductors
Cathodoluminescence
X-ray microanalysis
Resolution of the SEM
Environmental SEM
See also
References
External links
General
History
Images
Gallery of SEM images

Scanning process


Low-temperature SEM magnification series for a snow crystal.

In a typical SEM, electrons are thermionically emitted from a tungsten or lanthanum hexaboride (LaB6) cathode and are accelerated towards an anode; alternatively, electrons can be emitted via field emission (FE). Tungsten is used because it has the highest melting point and lowest vapour pressure of all metals, thereby allowing it to be heated for electron emission. The electron beam, which typically has an energy ranging from a few hundred eV to 100 keV, is focused by one or two condenser lenses into a beam with a very fine focal spot sized 0.4 nm to 5 nm. The beam passes through pairs of scanning coils or pairs of deflector plates in the electron optical column, typically in the objective lens, which deflect the beam horizontally and vertically so that it scans in a raster fashion over a rectangular area of the sample surface.
When the primary electron beam interacts with the sample, the electrons lose energy by repeated scattering and absorption within a teardrop-shaped volume of the specimen known as the interaction volume, which extends from less than 100 nm to around 5 µm into the surface. The size of the interaction volume depends on the electrons' landing energy, the atomic number of the specimen and the specimen's density. The energy exchange between the electron beam and the sample results in the emission of electrons and electromagnetic radiation, which can be detected to produce an image, as described below.
An insect coated in gold, having been prepared for viewing with a scanning electron microscope.

Sample preparation


As specimens are inundated by an influx of electrons, a charge will tend to accumulate unless the specimen is earthed. This is typically achieved by coating objects with a nanometre-thick layer of gold using a sputter coater, after mounting them on a metal stub or carbon film. Due to the loss of information such as colour and lustre, gold coating has until recently been considered a semi-destructive process, as it has not been possible to remove the coating without intensive handling, often damaging the specimen; a simple but novel technique involving potassium cyanide has recently been reported to permit such removal.[3]
Alternative techniques, for example the low-vacuum environmental SEM, allow samples to be imaged without such plating, but cannot render the high resolution available by the conventional method.

Detection of secondary electrons


The most common imaging mode monitors low energy (<50 eV) secondary electrons. Due to their low energy, these electrons originate within a few nanometers from the surface. The electrons are detected by an Everhart-Thornley detector which is a type of scintillator-photomultiplier device and the resulting signal is rendered into a two-dimensional intensity distribution that can be viewed and saved as a Digital image. This process relies on a raster-scanned primary beam. The brightness of the signal depends on the number of secondary electrons reaching the detector. If the beam enters the sample perpendicular to the surface, then the activated region is uniform about the axis of the beam and a certain number of electrons "escape" from within the sample. As the angle of incidence increases, the "escape" distance of one side of the beam will decrease, and more secondary electrons will be emitted. Thus steep surfaces and edges tend to be brighter than flat surfaces, which results in images with a well-defined, three-dimensional appearance. Using this technique, resolutions less than 1 nm are possible.

Detection of backscattered electrons


Backscattered electrons consist of high-energy electrons originating in the electron beam, that are reflected or back-scattered out of the specimen interaction volume. Backscattered electrons may be used to detect contrast between areas with different chemical compositions, especially when the average atomic number of the various regions is different, since the brightness of the BSE image tends to increase with the atomic number.
Backscattered electrons can also be used to form an 'electron backscatter diffraction' (EBSD) image. This image can be used to determine the crystallographic structure of the specimen.
There are fewer backscattered electrons emitted from a sample than secondary electrons. The number of backscattered electrons leaving the sample surface upward might be significantly lower than those that follow trajectories toward the sides. Additionally, in contrast with the case with secondary electrons, the collection efficiency of backscattered electrons cannot be significantly improved by a positive bias common on Everhart-Thornley detectors. This detector positioned on one side of the sample has low collection efficiency for backscattered electrons due to small acceptance angles. The use of a dedicated backscattered electron detector above the sample in a "doughnut" type arrangement, with the electron beam passing through the hole of the doughnut, greatly increases the solid angle of collection and allows for the detection of more backscattered electrons.

Beam-injection analysis of semiconductors


The nature of the SEM's probe, energetic electrons, makes it uniquely suited to examining the optical and electronic properties of semiconductor materials. The high-energy electrons from the SEM beam will inject charge carriers into the semiconductor. Thus, beam electrons lose energy by promoting electrons from the valence band into the conduction band, leaving behind holes.
In a direct bandgap material, recombination of these electron-hole pairs will result in cathodoluminescence; if the sample contains an internal electric field, such as is present at a p-n junction, the SEM beam injection of carriers will cause electron beam induced current (EBIC) to flow.
Cathodoluminescence and EBIC are referred to as "beam-injection" techniques, and are very powerful probes of the optoelectronic behavior of semiconductors, particularly for studying nanoscale features and defects.

Cathodoluminescence


Cathodoluminescence, the emission of light when atoms excited by high-energy electrons return to their ground state, is analogous to UV-induced fluorescence, and some materials such as zinc sulphide and some fluorescent dyes, exhibit both phenomena. Cathodoluminescence is most commonly experienced in everyday life as the light emission from the inner surface of the cathode ray tube in television sets and computer CRT monitors. In the SEM, CL detectors either collect all light emitted by the specimen, or can analyse the wavelengths emitted by the specimen and display a spectrum or an image of the cathodoluminescence in real colour.

X-ray microanalysis


X-rays, which are also produced by the interaction of electrons with the sample, may also be detected in an SEM equipped for energy-dispersive X-ray spectroscopy or wavelength dispersive X-ray spectroscopy.

Resolution of the SEM


The spatial resolution of the SEM depends on the size of the electron spot, which in turn depends on the magnetic electron-optical system which produces the scanning beam. The resolution is also limited by the size of the interaction volume, or the extent to which the material interacts with the electron beam. The spot size and the interaction volume both might be large compared to the distances between atoms, so the resolution of the SEM is not high enough to image individual atoms, as is possible in the transmission electron microscope (TEM). The SEM has compensating advantages, though, including the ability to image a comparatively large area of the specimen; the ability to image bulk materials (not just thin films or foils); and the variety of analytical modes available for measuring the composition and nature of the specimen. Depending on the instrument, the resolution can fall somewhere between less than 1 nm and 20 nm. In general, SEM images are easier to interpret than TEM images.

Environmental SEM


Conventional SEM requires samples to be imaged under vacuum, which means that samples that would produce a significant amount of vapour, e.g. biological samples, need to be either dried or cryogenically frozen. This means that processes involving transitions to or from liquids or gases, such as the drying of adhesives or melting of alloys, liquid transport, chemical reactions and solid-air-gas systems in general could not be observed.
The first commercial development of the Environmental SEM (ESEM) in the late 1980s
[4]
[5] allowed samples to be observed in low-pressure gaseous environments (e.g. 1-50 Torr) and high relative humidity (up to 100%). This was made possible by the development of a secondary-electron detector
[6]
[5] capable of operating in the presence of water vapour and by the use of pressure-limiting apertures with differential pumping in the path of the electron beam to separate the vacuum regions around the gun and lenses from the sample chamber.
The first commercial ESEMs were produced by the ElectroScan Corporation in USA in 1988 [8]. ElectroScan were later taken over by Philips (now FEI Company) in 1996 [9].
ESEM is especially useful for non-metallic and biological materials because coating with carbon or gold is unnecessary. Plastics and Elastomers can now be routinely examined, as can biological samples. Coating is irreversible, and may reduce the value of the results obtained. Thus very small details on the surface of the sample may be concealed by the coating, let alone that coating is done under vacuum, which drastically alters hydrated specimens.

See also



List of surface analysis methods

TEM

References


1. Das Elektronen-Rastermikroskop. Theoretishce Grundlagen, , Manfred, von Ardenne, Zeitschrift fur Physik, 1938
2. Das Elektronen-Rastermikroskop. Praktische Ausfurung, , Manfred, von Ardenne, Z. Techn. Phys., 1938
3. , , , , ,
4. Foundations of environmental scanning electron microscopy, , G,D, Danilatos, Advances in Electronics and Electron Physics, 1988
5.
6. Theory of the Gaseous Detector Device in the ESEM, , G,D, Danilatos, Advances in Electronics and Electron Physics, 1990
7.
8. History of Electron Microscopy, 1980s
9. History of Electron Microscopy 1990s

External links


General


Notes on the SEM Notes covering all aspects of the SEM
History


Microscopy History links from the University of Alabama Department of Biological Sciences

The Early History and Development of SEM from Cambridge University Engineering Department

Milestones in the History of Electron Microscopy from the Swiss Federal Institute of Zürich

Environmental Scanning Electron Microscope (ESEM) history
Images


Tescan Image Gallery Some great SEM images of various specimens, as well as analytical results

Dennis Kunkel Microscopy, Inc. Large collection of SEM images - mostly false colour

Jeol SEM Images Twelve SEM images of various specimens

SEM Lab at the Smithsonian National Museum of Natural History; includes a gallery of images

Rippel Electron Microscope Facility Many dozens of (mostly biological) SEM images from Dartmouth College.

Gallery of SEM images


The following are examples of images taken using a scanning electron microscope.


This article provided by Wikipedia. To edit the contents of this article, click here for original source.

psst.. try this: add to faves